Full credit to the source of this post http://www.leica-microsystems.com/science-lab/thin-metal-foils-with-coatings-sample-preparation-for-sem/

Application Note for Leica EM RES102 – Thin foils are mostly unstable because of their thickness of a few microns. This makes it difficult to do slope cutting without any protection of the sample. A common realisation to protect the sample surface is by sticking a cover glass on top of the sample. Another issue is cutting the foils before ion milling. The sample edge should be flat and sharp without any broken areas. A razor blade is mostly the best solution. A protected sample can salso be sawed with a wire saw.